Arca
Skip to content
Tools
Tools

Power supply provided with Afinion 2

Recall Z-2192-2020, part of FDA recall event 85313. See Abbott Diagnostics Technologies AS's company page.

Abbott Diagnostics Technologies AS started this Class II device recall on Mar 23, 2020 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Mar 23, 2020
  • Status: Ongoing

Why It Was Recalled

Class II

The analyzer system may have an electrostatic discharge and sparking of the power supply provided with the analyzers resulting in the power supply becoming non-functional.

Record

Product
Power supply provided with Afinion 2, Catalog numbers 1116553, 1116554, 1116556, 1116557, 1116597, 1116598, 1116663, 1116679, 1116680, 1116681, 1116682, 1116684, 1116770, 1116771, 1116772, 1116777, 1116778, 1116970, 1116971, 1116985, 1116986, IVD. The firm name on the label is Abbott Diagnostics Technologies AS, Oslo, Norway. Power supply provided with Alere Afinion AS100 Analyzer, Catalog numbers 1115175, 1115390, 1116049, 1116050, 1116053, 1116054, 1116456, and 1116980. The firm name on the label is Alere Technologies AS, Oslo, Norway.
Quantity
42,237 instruments
Lots and codes
All serial numbers are affected.
Distribution
Distribution was made to NJ. There was no direct government/military distribution. Foreign distribution was made to Canada, Argentina, Australia, Bangladesh, Brazil, Cambodia, Chile, China, Colombia, Denmark, Finland, France, Georgia, Germany, Guatemala, Hong Kong, Iceland, India, Italy, Ivory Coast, Japan, Korea, Madagascar, Malaysia, Mexico, Mongolia, Myanmar, The Netherlands, Norway, Oman, Pakistan, Peru, Philippines, Poland, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Timur, Trinidad & Tobago, United Kingdom, and Zimbabwe.
Recalling firm
Abbott Diagnostics Technologies AS, Oslo, N/A, Norway
How the firm notified
Letter
FDA classified
May 29, 2020

Abbott Diagnostics Technologies AS's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports