Arca
Skip to content
Tools
Tools

lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75…

Recall Z-2140-2018, part of FDA recall event 80096. See Alere San Diego's company page.

Alere San Diego Inc. DBA Immunalysis Corporation started this Class III device recall on Nov 28, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated Jan 29, 2021.

  • Classification: Class III
  • Started: Nov 28, 2017
  • Status: Terminated

Why It Was Recalled

Class III

The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.

Record

Product
lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.
Quantity
37 kits
Lots and codes
Part Number: 3006-5, Lot Number: EK16135
Distribution
US Distribution to the states of : CA, CO, CT, FL, LA, MD, NC, NJ, OH, OK, TN, TX, WI, and WV
Recalling firm
Alere San Diego Inc. DBA Immunalysis Corporation, Pomona, CA, United States
How the firm notified
Letter
FDA classified
Jun 8, 2018

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports