Arca
Skip to content
Tools
Tools

MED EL Cochlear Implant System

Recall Z-2137-2019, part of FDA recall event 81683. See MED-EL's company page.

MED-EL Elektromedizinische Gereate, Gmbh started this Class II device recall on Oct 16, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated May 22, 2020.

  • Classification: Class II
  • Started: Oct 16, 2018
  • Status: Terminated
  • Products in this event: 3

Why It Was Recalled

Class II

Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.

Record

Product
MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types: a. SONATAti100 FLEXsoft; b. SONATAti100 H Standard; c. SONATAti100 Standard Product Usage: The SONATATIi100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatible MED EL external components. The MED EL Cochlear Implant System is intended to evoke auditory sensations via electrical stimulation of the auditory pathways for severely to profoundly hearing impaired individuals who obtain little or no benefit from acoustic amplification in the best aided condition.
Quantity
10 devices
Lots and codes
a. SONATAti100 FLEXsoft, Serial Number 150489; b. SONATAti100 H Standard, Serial Numbers 146714, 164324; c. SONATAti100 Standard, Serial Numbers 145179, 217623, 144443, 145228, 158919, 147748, 223013.
Distribution
Worldwide distribution - US Nationwide in the states of California and Illinois. Countries of ARGENTINA, AUSTRALIA, AUSTRIA, BRAZIL, BULGARIA, CHINA, FRANCE, GEORGIA, KOREA, REPUBLIC OF, POLAND, RUSSIAN FEDERATION, SPAIN, SWEDEN.
Recalling firm
MED-EL Elektromedizinische Gereate, Gmbh, Innsbruck, N/A, Austria
How the firm notified
Letter
FDA classified
Aug 2, 2019

Other Products in This Recall

Recall event 81683

MED-EL Elektromedizinische Gereate, Gmbh's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.