Mobius3D Product Usage: Mobius3D software is used for quality assurance and treatment…
Recall Z-2100-2017, part of FDA recall event 77177. See Mobius Medical Systems's company page.
Mobius Medical Systems, LP started this Class II device recall on Feb 23, 2015 (voluntary: firm initiated); FDA lists it as terminated, terminated Jul 10, 2017.
- Classification: Class II
- Started: Feb 23, 2015
- Status: Terminated
Mobius3D version 1.5.0 contained a defect in software code which affects users who perform beam customization, and may lead to a discrepancy in dose calculation between this version of Mobius3D (1.5.0) and the prior version (1.4.2), where none should have been expected.
- Product
- Mobius3D Product Usage: Mobius3D software is used for quality assurance and treatment plan verification in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.
- Quantity
- 270
- Lots and codes
- Version 1.5.0 and 1.5.2 of the Mobius3D software is affected.
- Distribution
- Worldwide distribution - US Nationwide and the countries of Sweden, Canada, Netherlands, Denmark, Spain, Switzerland, Australia, Austria, Israel, United Kingdom, France
- Recalling firm
- Mobius Medical Systems, LP, Houston, TX, United States
- How the firm notified
- Letter
- FDA classified
- May 16, 2017