Cadence Size 5 Left Tibial Tray Implant
Recall Z-2086-2016, part of FDA recall event 74173. See Integra LifeSciences's company page.
Integra LifeSciences Corp. started this Class II device recall on Apr 29, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Oct 19, 2016.
- Classification: Class II
- Started: Apr 29, 2016
- Status: Terminated
As a result of an internal review of all labeling for the Cadence System, it was determined that the outer package label for the Size 5 Left Tibial Tray Implant reflects the material composition incorrectly. The labeling incorrectly reflects the material as Cobalt Chromium instead of Titanium.
- Product
- Cadence Size 5 Left Tibial Tray Implant, designed to treat ankle arthritis through replacement of the ankle joint with a prosthesis, thereby reducing pain, restoring alignment, and allowing for movement at the replaced joint. Catalog Number: 10207205; Lot Number #QJ0248
- Quantity
- 12 units
- Lots and codes
- Cat No. 10207205, Lot: QJ0248
- Distribution
- NC and PA
- Recalling firm
- Integra LifeSciences Corp., Plainsboro, NJ, United States
- How the firm notified
- FDA classified
- Jun 29, 2016
Newest first, one product per recall event; search them all
- AURORA Surgiscope System, Sterile, single use device that contains a Sheath, Obturator, and Imager. Manufacturer's…Z-1295-2025 · Class II · Feb 6, 2025 · Ongoing
- Brand Name: MediHoney¿ Gel Product Name: MediHoney¿ Wound & Burn Dressing Gel Model/Catalog Number: 31815 Software…Z-1168-2025 · Class II · Jan 10, 2025 · Ongoing
- CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810PL. Implantable device that…Z-0978-2025 · Class II · Dec 16, 2024 · Ongoing
- Duraform Dural Graft Implant, Size 3x3 , Model Number 80-1477US Lot CT007694Z-0262-2025 · Class II · Sep 26, 2024 · Ongoing
- SURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated for the…Z-0240-2025 · Class II · Sep 20, 2024 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports