Arca
Skip to content
Tools
Tools

K-wire, Smooth Double-pointed 0.8 x 90mm

Recall Z-2084-2016, part of FDA recall event 74224. See Integra LifeSciences's company page.

Integra LifeSciences Corp. started this Class II device recall on May 20, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Nov 16, 2016.

  • Classification: Class II
  • Started: May 20, 2016
  • Status: Terminated

Why It Was Recalled

Class II

The affected K-wires do not meet specification requirements. Specifically, a) K-wires were made of 316LVM stainless steel instead of 304 stainless steel. b) K-wires have passivated surface finish instead of electro-polish surface finish. c) K-wires have incorrect length and diameter.

Record

Product
K-wire, Smooth Double-pointed 0.8 x 90mm, a component of the METASURG Capture Screw System The K-wire is used in conjunction with the 2.0 mm cannulated screws for the METASURG Capture Screw System. The K-wire is used by the physician to insert into the bone and to act as a guide for drilling, countersinking and installing screws
Quantity
182 units
Lots and codes
Lot 237830TE
Distribution
Nationwide Distribution to OK, CA, TX, WA, GA, NY, MS, NY, PA
Recalling firm
Integra LifeSciences Corp., Plainsboro, NJ, United States
How the firm notified
Letter
FDA classified
Jun 28, 2016

Integra LifeSciences Corp.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc