ProCair Mattress: (a) PROCAIR PLUS 42X80
Recall Z-2034-2018, part of FDA recall event 80113. See Joerns Healthcare's company page.
Joerns Healthcare started this Class II device recall on May 3, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated Sep 28, 2023.
- Classification: Class II
- Started: May 3, 2018
- Status: Terminated
- Products in this event: 2
Joerns Healthcare has identified a potential issue with some of the P.R.O. Matt and ProCair Mattresses. For the identified serial number range of P.R.O. Matt and ProCair Mattresses incorrect connectors (part #s 206-0401 & 206-0405) were provided by a Joerns supplier without the required internal shutoff valves, these shutoff valves prevent air from escaping from the mattress air cells.
- Product
- ProCair Mattress: (a) PROCAIR PLUS 42X80, MATERIAL NUMBER 222-4280 (b) PROCAIR PLUS 36X80, MATERIAL NUMBER 222-3680 (c) PROCAIR PLUS 84" MATTRESS, MATERIAL NUMBER 222-6384 Product Usage: Non powered flotation therapy mattress. The device is intended to treat or prevent decubitus ulcers (bed sores).
- Quantity
- 46 units
- Lots and codes
- (a) PROCAIR PLUS 42X80, MATERIAL NUMBER 222-4280 Serial Numbers: 1000054866 1000054860 1000051368 1000054859 1000054855 1000054857 (b) PROCAIR PLUS 36X80, MATERIAL NUMBER 222-3680 Serial Numbers: 1000053319 1000053312 1000052886 1000053325 1000053323 1000052420 1000052259 1000052426 1000052425 1000052432 1000052891 1000052458 1000052429 1000052256 1000052888 1000053990 1000051418 1000051419 1000052457 1000052252 1000051411 1000055556 1000052460 1000054129 1000053979 1000049129 1000051412 1000052424 1000052455 1000054117 1000053982 1000053983 1000054122 1000054121 (c) PROCAIR PLUS 84"…
- Distribution
- Worldwide Distribution - US Nationwide in the states of UK, Canada, Holland
- Recalling firm
- Joerns Healthcare, Arlington, TX, United States
- How the firm notified
- Letter
- FDA classified
- May 30, 2018
- P.R.O. Matt: (a) PROMATT PLUS 36 X 76 MATRESS, NYLON, MATERIAL NUMBER PMP3676NZBA (b) PROMATT PLUS 84" MATRESS,…Z-2033-2018 · Class II · May 3, 2018 · Terminated
- Joerns EasyCare Bed Product Usage: The AC-powered adjustable hospital bed is a device intended for medical…Z-0151-2015 · Class II · Sep 24, 2014 · Terminated