Arca
Skip to content
Tools
Tools

TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The…

Recall Z-1965-2016, part of FDA recall event 74114. See Integra LifeSciences's company page.

Integra LifeSciences Corp. started this Class II device recall on May 10, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Dec 16, 2016.

  • Classification: Class II
  • Started: May 10, 2016
  • Status: Terminated

Why It Was Recalled

Class II

As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior lot numbers manufactured before Integra acquired the product line reflect an incorrect number for the first mm depth marking on the gauge. The number should be 4, it is 6. 3) The space / holder

Record

Product
TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is used by the physician to determine the correct screw length when implanting a 2.0 or 2.5mm TS Series screw for the Ti6 Internal Fixation System.
Quantity
71 units
Lots and codes
Lot UP15H2, Lot UP09D3, Lot UP07E2
Distribution
US Distribution to states of: CO, FL, IL, MI, NJ, NY, SC, TX, and VA.
Recalling firm
Integra LifeSciences Corp., Plainsboro, NJ, United States
How the firm notified
Letter
FDA classified
Jun 14, 2016