NIPRO Safe Touch Safety Scalp Vein Set
Recall Z-1834-2014, part of FDA recall event 65274. See Nipro Medical's company page.
Nipro Medical Corporation started this Class II device recall on May 4, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Mar 2, 2015.
- Classification: Class II
- Started: May 4, 2013
- Status: Terminated
There is a possibility of a crack in the CP luer connector. The position of the crack is at the connection site of the CP luer connector and the tubing. The crack could cause or contribute to leakage of fluids.
- Product
- NIPRO Safe Touch Safety Scalp Vein Set
- Quantity
- 176,000 pieces
- Lots and codes
- Product Code: PR+25G19, Lots #s 12H31H and 12I14H; Product Code: SPR+21G19, Lots #s 12I24 and 12J11; Product Code: PR-25G19, Lot # 12H27H; and PR-23G19, Lot # 12I12H.
- Distribution
- Worldwide Distribution - USA and the countries of Canada, Bolivia, Chile, Uruguay, Belize, Dominica, Honduras, Jamaica, Paraguay and Venezuela.
- Recalling firm
- Nipro Medical Corporation, Miami, FL, United States
- How the firm notified
- Letter
- FDA classified
- Jun 19, 2014
Newest first, one product per recall event; search them all
- Hemodialysis System, Surdial DX REF: MC+SDX01 Rx OnlyZ-0912-2026 · Class II · Oct 27, 2025 · Ongoing
- Nipro Blood Tubing set with Priming Set and Transducer Protectors. Each device is packaged in a film pouch with 24…Z-1876-2018 · Class II · Dec 19, 2017 · Terminated
- Nipro Safe Touch TULIP Safety Fistula Needle, FT+152530TP, for use as a blood access device for blood purification and…Z-0161-2016 · Class II · Aug 13, 2015 · Terminated
- NIPRO SET / SLIMLINE BLOOD TUBING SET for Hemodialysis with Transducer Protectors and Priming Set A disposable…Z-0624-2016 · Class II · Mar 30, 2015 · Terminated
- NIPRO SafeTouch II Safety AVF Needle, 17G x 1 ¿ inch. Item code FS+173230BC.Z-0599-2016 · Class III · Apr 1, 2014 · Terminated
- AquaLiner Hydrophilic Guidewire 0.035" X 260cm, Straight/ Regular. AquaLiner Hydrophilic Guidewire is designed for…Z-0136-2013 · Class II · Jul 27, 2012 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports