Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea…
Recall Z-1789-2014, part of FDA recall event 68226. See Children's Medical Ventures's company page.
Childrens Medical Ventures started this Class II device recall on Apr 29, 2014 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 6, 2015.
- Classification: Class II
- Started: Apr 29, 2014
- Status: Terminated
The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.
- Product
- Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.
- Quantity
- 112
- Lots and codes
- 989805603611 / 1014557, 989805603631 / 1014559, 989805604511 / 1028969, 989805604521 / 1028970, 989805604531 / 1028971, 989805604631 / 1030271, 989805605051 / 1038140, 989805605061 / 1038141, 989805608801 / U1014557, 989805608811 / U1014559, 989805608831 / U1028969, 989805608841 / U1028970, 989805608851 / U1038140, 989805604231 / 1023384
- Distribution
- Worldwide Distribution - USA including the states of MI, WI, SC, FL, TX, KS, MA, AZ, UT, WV, IN, MN, NV and VA., and the countries of Australia, Belgium and Canada.
- Recalling firm
- Childrens Medical Ventures, Monroeville, PA, United States
- How the firm notified
- Letter
- FDA classified
- Jun 13, 2014
- Gel-E Donut, Squishon 2 gel pillows Product Usage:: The products are intended to support and cradle a baby s head…Z-2192-2014 · Class I · May 28, 2014 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports