Arca
Skip to content
Tools
Tools

Temperature Sensing 100% Silicone Foley Catheter

Recall Z-1782-2020, part of FDA recall event 85327. See Degania Medical Devices Pvt.'s company page.

Degania Medical Devices Pvt. Ltd. started this Class II device recall on Mar 17, 2020 (voluntary: firm initiated); FDA lists it as terminated, terminated Oct 25, 2024.

  • Classification: Class II
  • Started: Mar 17, 2020
  • Status: Terminated
  • Products in this event: 11

Why It Was Recalled

Class II

Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.

Record

Product
Temperature Sensing 100% Silicone Foley Catheter, Part Number 102201101463MD - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.
Quantity
16393
Lots and codes
Lot Numbers: V18022033 V18025110 V18025111 V18029225 V18029226 V19024273 V19013327 V19025875 V19016720 V19025874
Distribution
Worldwide distribution - US Nationwide distribution including state of IL and the country of Mexico.
Recalling firm
Degania Medical Devices Pvt. Ltd., Gurgaon, N/A, India
How the firm notified
Letter
FDA classified
Apr 23, 2020

Other Products in This Recall

Recall event 85327

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc