Arca
Skip to content
Tools
Tools

Skytron Integrity 270 Steam Sterilizer and Skytron Integrity 270VP Steam Product Usage:…

Recall Z-1779-2019, part of FDA recall event 82650. See Skytron's company page.

Skytron, Div. The KMW Group, Inc started this Class II device recall on Apr 18, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 17, 2020.

  • Classification: Class II
  • Started: Apr 18, 2019
  • Status: Terminated

Why It Was Recalled

Class II

Weld attaching the air heater assembly to the main sterilizer unit has shown signs of fatigue/failure

Record

Product
Skytron Integrity 270 Steam Sterilizer and Skytron Integrity 270VP Steam Product Usage: The SKYTRON Integrity 270 Steam Sterilizer for use in health care facilities. The Integrity Sterilizer incorporates high-pressure steam to sterilize nonporous or porous items and materials used in the health care industry
Quantity
5 units
Lots and codes
Serial Numbers: 14092619, 14092620, 14102621, 17123520, 18113521
Distribution
US Nationwide Distribution - AK, IA, MS
Recalling firm
Skytron, Div. The KMW Group, Inc, Grand Rapids, MI, United States
How the firm notified
Letter
FDA classified
Jun 13, 2019

Skytron, Div. The KMW Group, Inc's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports