Ascom Mobile Monitoring Gateway (versions 4.1.1 and 4.2.0) and Ascom Unite Connect for…
Recall Z-1638-2016, part of FDA recall event 73578. See Ascom (US)'s company page.
Ascom US, INC. started this Class II device recall on Dec 10, 2015 (voluntary: firm initiated); FDA lists it as terminated, terminated Oct 13, 2017.
- Classification: Class II
- Started: Dec 10, 2015
- Status: Terminated
Due to a malfunction of software, the secondary module will not automatically take over messaging if the primary module fails after a software upgrade to one of the other affected versions. Messages from GE Carescape/Nurse Call will not forward to handset until primary module is restored.
- Product
- Ascom Mobile Monitoring Gateway (versions 4.1.1 and 4.2.0) and Ascom Unite Connect for Nurse Call (version 2.2.0).
- Quantity
- 319
- Lots and codes
- Ascom Mobile Monitoring Gateway version 4.1.1 & 4.2.0 and Ascom Unite Connect for Nurse Call version 2.2.0
- Distribution
- CA, CT, FL, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MO, NJ, NM, OH, OK, PA, TN, TX, VA, WA, WI and Hawaii. Australia, Austria, Belgium, Canada, France, Germany, Netherlands, Poland, Qatar, Sweden, Switzerland, and United Arab Emirates.
- Recalling firm
- Ascom US, INC., Morrisville, NC, United States
- How the firm notified
- Letter
- FDA classified
- May 10, 2016
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports