Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name:…
Recall Z-1634-2026, part of FDA recall event 98402. See AMO Puerto Rico Manufacturing's company page.
AMO Puerto Rico Manufacturing, Inc. started this Class II device recall on Feb 19, 2026 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Feb 19, 2026
- Status: Ongoing
Due to identifying a limited quantity of intraocular lenses that may have a haptic that sticks to the optic which prevents the lens from unfolding as consistently as expected.
- Product
- Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Model/Catalog Number: DIB00 Software Version: N/A Product Description: The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. The lens is preloaded and pre-assembled in the delivery system. The TECNIS SIMPLICITY" Delivery System Model DIB00 contains the TECNISEyhance" IOL, which is a one-piece, foldable, posterior chamber lens with an overall diameter of 13.0 mm and an optic diameter of 6.0 mm. The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. Component: No
- Quantity
- 361
- Lots and codes
- Lot Code: GTIN: 05050474731776 SKU: DIB00U0210-12 Serial Numbers: 2409342403 2410072403 2409612403 2409692403 2409752403 2891682403 2891132403 2892032403 2531882403 2531902403 2531212403 2530842403 2531242403 2531892403 2531622403
- Distribution
- U.S Nationwide distribution in the states of CA,FL, IL, ME,MS, NJ, NY, OK, OR, TX, and WI.
- Recalling firm
- AMO Puerto Rico Manufacturing, Inc., Anasco, PR, United States
- How the firm notified
- Letter
- FDA classified
- Mar 24, 2026
Newest first, one product per recall event; search them all
- AMO TECNIS CL Foldable Silicone IOL with OptiEdge Design, Model Z9002. Indicated for primary implantation for the…Z-2272-2014 · Class II · May 13, 2014 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports