Arca
Skip to content
Tools
Tools

Merits E604 Pilot Navigator Stairlift - Product Usage: To transport patients up stairs.

Recall Z-1632-2020, part of FDA recall event 85154. See Merits Holdings Co. dba Merits Health Products's company page.

Merits Holdings Co. dba Merits Health Products, Inc. started this Class II device recall on Mar 2, 2020 (voluntary: firm initiated); FDA lists it as terminated, terminated Sep 18, 2023.

  • Classification: Class II
  • Started: Mar 2, 2020
  • Status: Terminated

Why It Was Recalled

Class II

Pinion gear may disengage from track allowing lift to slip or jolt.

Record

Product
Merits E604 Pilot Navigator Stairlift - Product Usage: To transport patients up stairs.
Quantity
82 units.
Lots and codes
Model: E60424YLMUB; SN: MU180600588; (01) 04714379292589 (11)190604 (21) MU190600025.
Distribution
US Nationwide distribution in the states of AZ, CA, CT, FL, GA, IN, MA, MD, ME, MO, NC, OH, PA, TX, and VA.
Recalling firm
Merits Holdings Co. dba Merits Health Products, Inc., Fort Myers, FL, United States
How the firm notified
Letter
FDA classified
Apr 1, 2020