Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal…
Recall Z-1631-2026, part of FDA recall event 98556. See Koven Technology's company page.
Koven Technology, Inc. started this Class II device recall on Mar 4, 2026 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Mar 4, 2026
- Status: Ongoing
Product labeling includes a fetal indication for use that is not cleared under its 510(k).
- Product
- Brand Name: BiDop 3 Product Name: BiDop 3 Pocket Doppler (ES-100V3) with BT2M20S8C fetal probe Model/Catalog Number: ES-100V3 (device); BT2M20S8C (2 MHz fetal probe). Software Version: N/A Product Description: Hand-held, battery-powered ultrasonic Doppler blood-flow detector with LCD numeric display, used with interchangeable probes to detect vascular blood-flow signals and display heart rate; affected units were packaged with a 2 MHz BT2M20S8C fetal probe and labeled for both vascular and fetal applications.
- Quantity
- 34
- Lots and codes
- Lot Code: BIDOP 3 #00894912002043 BT2M20S8C #00894912002326
- Distribution
- US Nationwide distribution in the states of LA, NC, MT, TX MA, OR, NY, IL, CO, GA, FL, WA, WI, MO, AZ, IN.
- Recalling firm
- Koven Technology, Inc., Saint Louis, MO, United States
- How the firm notified
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- FDA classified
- Mar 24, 2026
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports