Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures…

Recall Z-1617-2023, part of FDA recall event 92136. See Omnicell's company page.

OmniCell, Inc. started this Class II device recall on Mar 27, 2023 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Mar 27, 2023
  • Status: Ongoing

Why It Was Recalled

Class II

IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling

Record

Product
Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures and syringes Model Number: 298921373
Quantity
6 units
Lots and codes
UDI-DI: 00860006285005 Serial Numbers: IVXS000014, IVXS000015, IVXS000016, IVXS000017, IVXS000018. IVXS000019
Distribution
AL, NY
Recalling firm
OmniCell, Inc., Warrendale, PA, United States
How the firm notified
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
FDA classified
May 22, 2023