Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x…
Recall Z-1552-2020, part of FDA recall event 84698. See Quidel Cardiovascular's company page.
QUIDEL CARDIOVASCULAR INC started this Class II device recall on Apr 9, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Jan 14, 2021.
- Classification: Class II
- Started: Apr 9, 2019
- Status: Terminated
Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.
- Product
- Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0.25 mL vials per box. Quidel Cardiovascular Inc. 9975 Summers Ridge Road San Diego, CA 92121 USA Quidel.com
- Quantity
- 123 kits
- Lots and codes
- Lot Number: C3470AN Part Number 94413
- Distribution
- US: ID, FL, GA, IL, TX, KY, MO, AL, UT, IA, WA, OH, AZ, NY, OK, OR, IN, KS Non-US: CA, NL
- Recalling firm
- QUIDEL CARDIOVASCULAR INC, San Diego, CA, United States
- How the firm notified
- Letter
- FDA classified
- Mar 20, 2020