Arca
Skip to content
Tools
Tools

Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x…

Recall Z-1552-2020, part of FDA recall event 84698. See Quidel Cardiovascular's company page.

QUIDEL CARDIOVASCULAR INC started this Class II device recall on Apr 9, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Jan 14, 2021.

  • Classification: Class II
  • Started: Apr 9, 2019
  • Status: Terminated

Why It Was Recalled

Class II

Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.

Record

Product
Quidel Triage TOX Drug Screen Control 1 Kit Box containing 5 Control 1 vials with 5 x 0.25 mL vials per box. Quidel Cardiovascular Inc. 9975 Summers Ridge Road San Diego, CA 92121 USA Quidel.com
Quantity
123 kits
Lots and codes
Lot Number: C3470AN Part Number 94413
Distribution
US: ID, FL, GA, IL, TX, KY, MO, AL, UT, IA, WA, OH, AZ, NY, OK, OR, IN, KS Non-US: CA, NL
Recalling firm
QUIDEL CARDIOVASCULAR INC, San Diego, CA, United States
How the firm notified
Letter
FDA classified
Mar 20, 2020

QUIDEL CARDIOVASCULAR INC's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports