TX1 Tissue Removal System. TX1 Procedure Pack P/N 554-1003-001 (outer box)
Recall Z-1552-2014, part of FDA recall event 68074. See Tenex Health's company page.
Tenex Health Inc started this Class II device recall on Apr 14, 2014 (voluntary: firm initiated); FDA lists it as terminated, terminated Jul 16, 2014.
- Classification: Class II
- Started: Apr 14, 2014
- Status: Terminated
The sterile barrier in the packaging may be compromised due to cracks in the thermoform tray that may occur during transport of the product in extreme cold weather conditions, in conjunction with mechanical impact.
- Product
- TX1 Tissue Removal System. TX1 Procedure Pack P/N 554-1003-001 (outer box), TX1 MicroTip P/N 554-1002-001 (inner tray). For use with the TX System Console as an ultrasonic surgical aspirator of soft tissue.
- Quantity
- 9,927 units
- Lots and codes
- Lot Numbers: 00614-03, 06513-02, 24913-01, 00713-06, 06514-02, 25313-01, 00814-02, 06914-01, 25513-02, 01114-02, 07213-04, 25613-01, 01514-03, 07713-18, 26113-06, 01714-04, 08113-04, 26313-04, 01813-02, 08413-13, 26713-04, 02314-08, 09313-05, 26813-01, 02413-04, 10213-07, 27013-10, 02714-02, 10613-01, 27613-02, 02914-01, 11213-07, 28013-05, 03014-02, 11613-06, 28313-02, 03113-02, 12113-02, 29513-03, 03114-02, 12713-03, 29713-01, 03414-03, 13413-01, 30913-07, 03514-05, 14413-05, 31113-08, 03714-03, 15013-04, 31513-02, 03814-06, 15413-02, 31713-05, 04114-02, 15413-03, 32213-01, 04213-01,…
- Distribution
- Worldwide Distribution - USA (Nationwide) and Internationally to Panama and Singapore.
- Recalling firm
- Tenex Health Inc, Lake Forest, CA, United States
- How the firm notified
- Letter
- FDA classified
- May 6, 2014
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports