Joseph Lamelas Atrial Lift System
Recall Z-1538-2014, part of FDA recall event 68093. See Biorep Technologies's company page.
Biorep Technologies started this Class II device recall on Mar 19, 2014 (voluntary: firm initiated); FDA lists it as terminated, terminated Aug 7, 2014.
- Classification: Class II
- Started: Mar 19, 2014
- Status: Terminated
The support arm attached with an adhesive to the body of the blade is detaching while attempting to engage the visor component.
- Product
- Joseph Lamelas Atrial Lift System, Model Number MI-ALS-001. A single surgical instrument or tool for retraction of the atrial wall tissue in cardiac surgery.
- Quantity
- 80 boxes (480 units)
- Lots and codes
- Lot numbers: 166934413 (expires 2016/12/10), and 166903414 (expires 2017/02/03).
- Distribution
- US Distribution to a distributor in MN.
- Recalling firm
- Biorep Technologies, Miami, FL, United States
- How the firm notified
- Letter
- FDA classified
- May 2, 2014
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports