LipiFlow Thermal Pulsation System/LipiFlow

Recall Z-1535-2017, part of FDA recall event 76365. See Tearscience's company page.

Tearscience, Inc started this Class II device recall on Jan 30, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated Nov 27, 2017.

  • Classification: Class II
  • Started: Jan 30, 2017
  • Status: Terminated

Why It Was Recalled

Class II

Labeling: The labeled storage temperature is not consistent with the temperature conditions tested for the entire four year shelf life duration.

Record

Product
LipiFlow Thermal Pulsation System/LipiFlow, System Activator (Disposable), Rx Only, Sterile
Quantity
94,095 units.
Lots and codes
Model number LFD-1000 and LFD-1100, All lots
Distribution
US (nationwide)
Recalling firm
Tearscience, Inc, Morrisville, NC, United States
How the firm notified
Letter
FDA classified
Mar 22, 2017

Tearscience, Inc's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports