ThermaCare HEATWRAPS
Recall Z-1496-2019, part of FDA recall event 82781. See PF Consumer Healthcare 1's company page.
PF Consumer Healthcare 1 LLC started this Class II device recall on Apr 26, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Jun 11, 2024.
- Classification: Class II
- Started: Apr 26, 2019
- Status: Terminated
- Products in this event: 2
There is the potential that a HeatWrap could include cells that have a higher cell temperature than specified in lot S97473. The use of heat cell wrap with increased temperatures poses a potential risk of skin injuries such as burns/blisters and/or skin irritation on the wrap-applied area.
- Product
- ThermaCare HEATWRAPS, ADVANCED BACK PAIN THERAPY, SKU: a) F00573301020A b) F00573301026X c) F00573301009B Product Usage: ThermaCare(R) HeatWraps are indicated in adults to provide heat therapy for the temporary relief of minor muscular and joint pains associated with overexertion, strains, sprains, and arthritis and menstrual pain. ThermaCare HeatWraps are made up of heat cells enclosed within wraps that consists of a mixture of ingredients including soft iron, carbon or absorbent gelling material and brine solution (water, salt, and sodium thiosulphate). The heat cells get activated in the presence of air (oxygen) to produce a low-level heat.
- Quantity
- 225617 individual heat wraps
- Lots and codes
- Lot Numbers: a) S23902 b) W37940 c) S00639
- Distribution
- Worldwide distribution - US Nationwide and PR, and countries of Ireland, UK, Malta, Netherlands, Germany, and Switzerland.
- Recalling firm
- PF Consumer Healthcare 1 LLC, Albany, GA, United States
- How the firm notified
- Letter
- FDA classified
- May 28, 2019
- ThermaCare HEATWRAPS, ADVANCED BACK PAIN THERAPY, UPC 305733010037 - Product Usage: ThermaCare(R) HeatWraps are…Z-1495-2019 · Class II · Apr 26, 2019 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports