Arca
Skip to content
Tools
Tools

NaviNetics Skull Anchor Key and Drill Kit

Recall Z-1478-2024, part of FDA recall event 94163. See Navinetics's company page.

Navinetics Inc started this Class II device recall on Feb 26, 2024 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Feb 26, 2024
  • Status: Ongoing

Why It Was Recalled

Class II

Defective Skull Anchor Key. After attachment of the single use Skull Anchor Key to the patient, the user was unable to attach the CT Localizer to perform stereotactic Imaging. Attachment of the localizer is typically accomplished by fully screwing in the four thumbscrews of the localizer into the threaded holes found on the top surface of the Skull Anchor Key.

Record

Product
NaviNetics Skull Anchor Key and Drill Kit, REF NN1215
Quantity
20 kits
Lots and codes
UDI/DI 10850023620276, Lot Number 331401724
Distribution
US Nationwide distribution in the states of FL, MN, CO.
Recalling firm
Navinetics Inc, Rochester, MN, United States
How the firm notified
E-Mail
FDA classified
Apr 4, 2024

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports