Arca
Skip to content
Tools
Tools

The humeral stems trials are packaged in a kit for Titan Modular Shoulder System under…

Recall Z-1463-2018, part of FDA recall event 79597. See Integra LifeSciences's company page.

Integra LifeSciences Corp. started this Class II device recall on Dec 20, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated Feb 24, 2020.

  • Classification: Class II
  • Started: Dec 20, 2017
  • Status: Terminated

Why It Was Recalled

Class II

Incidents of stem trial breakage were reported to the firm suggesting that these fractures all occurred during insertion/impaction or extraction of the humeral stem trial whiling preparing the humeral canal and/or trialing. All cases resulted in a delay in surgery, with a variance of medical intervention required.

Record

Product
The humeral stems trials are packaged in a kit for Titan Modular Shoulder System under the following labeling: Catalog Number Product Description TRL092002506 Humeral Stem Trial, 6mm TRL092002507 Humeral Stem Trial, 7mm TRL092002508 Humeral Stem Trial, 8mm TRL092002509 Humeral Stem Trial, 9mm TRL092002510 Humeral Stem Trial, 10mm TRL092002511 Humeral Stem Trial, 11mm TRL092002512 Humeral Stem Trial, 12mm TRL092002513 Humeral Stem Trial, 13mm TRL092002514 Humeral Stem Trial, 14mm TRL092002515 Humeral Stem Trial, 15mm TRL092002516 Humeral Stem Trial, 16mm
Quantity
1619 units
Lots and codes
TRL092002506 Lot Numbers: 19419-1 15916-1 15457-1 201274-1 201893-1 2046177 TRL092002507 Lot Numbers: 19419-2 15916-2 15457-2 201274-2 201893-2 2046178 TRL092002508 Lot Numbers: 19419-3 15916-3 15457-3 201274-3 201893-3 2046179 TRL092002509 Lot Numbers: 19419-4 15916-4 15457-4 201274-4 201893-4 20461710 TRL092002510 Lot Numbers: 19419-5 15916-5 15457-5 201274-5 201893-5 20461711 TRL092002511 Lot Numbers: 19419-6 15916-6 15457-6 201274-6 201893-6 20461712 TRL092002512 Lot Numbers: 19419-7 15916-7 15457-7 201274-7 201893-7 20461713 TRL092002513 Lot Numbers: 19419-8 15916-8 15457-8 201274-8…
Distribution
Worldwide Distribution - USA (nationwide) Distribution.
Recalling firm
Integra LifeSciences Corp., Plainsboro, NJ, United States
How the firm notified
Letter
FDA classified
Apr 20, 2018

Integra LifeSciences Corp.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.