XTRAC Excimer Laser, Model No. AL8000
Recall Z-1455-2017, part of FDA recall event 76697. See Strata Skin Sciences's company page.
Strata Skin Sciences, Inc. started this Class II device recall on Jan 23, 2017 (fda mandated); FDA lists it as terminated, terminated Jun 5, 2017.
- Classification: Class II
- Started: Jan 23, 2017
- Status: Terminated
When the Liquid Guide ( LLG) is not used according to the operator manual, the switch control wires may become loose and shorted. Shorted wires may lead to unintentional laser firing.
- Product
- XTRAC Excimer Laser, Model No. AL8000, AL10000 The intended use is targeted UVB phototherapy for treatment of the skin conditions including psoriasis, vitiligo, atopic dermatitis, and leukoderma.
- Quantity
- US - 1000 , Foreign - 612
- Lots and codes
- SN 80007 80009 80010 80011 80018 80019 80020 80025 80031 80037 80038 80042 80043 80047 80061 80069 80073 80074 80077 80084 80092 80094 80097 80099 80102 80104 80113 80115 80116 80119 80121 80122 80124 80125 80128 80131 80134 80150 80151 80156 80158 80160 80161 80164 80168 80169 80174 80176 80177 80183 80186 80187 80193 80195 80199 80203 80208 80219 80221 80224 80225 80232 80238 80248 80249 80256 80257 80258 80259 80261 80278 80280 80293 80294 80298 80304 80310 80312 80314 80315 80317 80326 80327 80334 80336 80339 80348 80351 80352 80355 80358 80359 80360 80366 80369 80370 80372 80378 80379…
- Distribution
- USA (nationwide) Distribution
- Recalling firm
- Strata Skin Sciences, Inc., Carlsbad, CA, United States
- How the firm notified
- Letter
- FDA classified
- Mar 27, 2017