Teleflex MEDICAL HUDSON RCI One Way Valve w. Capped Monitoring Port
Recall Z-1403-2018, part of FDA recall event 79670. See Teleflex Medical Europe's company page.
Teleflex Medical Europe Ltd started this Class II device recall on Mar 27, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated May 4, 2021.
- Classification: Class II
- Started: Mar 27, 2018
- Status: Terminated
These one-way valves may disconnect at the joint between the two components that make up the device, which would cause an interruption of ventilation to the patient.
- Product
- Teleflex MEDICAL HUDSON RCI One Way Valve w. Capped Monitoring Port, REF 1644, QTY 50 Product Usage: for use in respiratory and anesthesia circuits to connect two or more components of a breathing system.
- Quantity
- 103910 units
- Lots and codes
- Lot Numbers: 74D1501706 74K1502092 74F1600367 74J1600117 74L1601916 74C1701733 74F1700707 74D1501731 74K1502045 74F1601072 74J1601101 74M1600284 74C1702365 74F1701866 74G1500664 74M1501145 74F1601874 74J1601563 74M1601685 74D1700030 74F1702612 74G1500663 74A1601936 74F1602470 74J1602238 74A1700107 74D1700363 74G1700034 74G1502412 74B1600334 74G1600466 74K1600148 74A1701335 74D1701038 74G1700488 74H1500466 74C1600162 74H1600217 74K1600287 74A1701403 74D1701507 74G1701144 74H1500189 74D1601420 74H1600516 74K1601020 74A1702060 74E1700450 74G1702002 74H1501260 74D1601965 74H1601371 74K1601924…
- Distribution
- US Nationwide in the states of AU, CA, NZ
- Recalling firm
- Teleflex Medical Europe Ltd, Athlone, Ireland
- How the firm notified
- Letter
- FDA classified
- Apr 17, 2018
Newest first, one product per recall event; search them all
- RUSCH Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF numbers: a) 504550 b) 504555 c) 504560 …Z-2320-2021 · Class II · Jul 16, 2021 · Terminated
- Teleflex MEDICAL Pilling(R) UNIVERSAL SHEARS 7 1/2", REF 142542 Product Usage: general surgical scissorsZ-2287-2018 · Class II · Jun 12, 2018 · Terminated
- Teleflex MEDICAL LMA(TM) MADgic(TM) LARYNGO-TRACHEAL MUCOSAL ATOMIZATION DEVICE, MAD700, QTY: 1 Intended for the…Z-1628-2018 · Class II · Mar 19, 2018 · Terminated
- Baker Jejunostomy Tubes: Product Code Equivalent Code (a) 655300160 655316 (b) 655500200 655520Z-1370-2018 · Class II · Mar 15, 2018 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports