Arca
Skip to content
Tools
Tools

VERICIS, Merge Cardio

Recall Z-1399-2025, part of FDA recall event 96344. See Merge Healthcare's company page.

Merge Healthcare, Inc. started this Class II device recall on Feb 19, 2025 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Feb 19, 2025
  • Status: Ongoing
  • Products in this event: 5

Why It Was Recalled

Class II

When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

Record

Product
VERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0, 10.2 P2 , 10.3; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
Quantity
11
Lots and codes
Software Versions and Part Numbers: 10.0: Part Numbers: 88-00527-00, 88-00528-00, 88-00529-00, 88-00530-00, 88-00531-00, 88-00532-00, 88-00533-00, 70-00008-00; 10.1.1.1: Part Numbers: 88-01094-00; 10.1.2: Part Numbers: 88-01074-00, 88-01075-00, 88-01076-00, 88-01077-00, 88-01078-00, 88-01079-00, 88-00781-00, 88-00803-00, 88-01080-00, 70-00473-00; 10.2.0: Part Numbers: 88-01115-00, 88-01116-00, 88-01117-00, 88-01118-00, 88-01119-00, 88-01120-00, 88-01121-00, 88-01122-00, 88-01123-00, 70-00491-00, 10.2 P2: Part Numbers: 88-01203-00 10.3 P1 UDI - (01)00842000100041(10)10.3.0.2009(11)170309
Distribution
US
Recalling firm
Merge Healthcare, Inc., Hartland, WI, United States
How the firm notified
E-Mail
FDA classified
Mar 20, 2025

Other Products in This Recall

Recall event 96344

Merge Healthcare, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports