CereLink ICP Monitor; Catalog No. 826820
Recall Z-1395-2022, part of FDA recall event 90457. See Integra LifeSciences's company page.
Integra LifeSciences Corp. started this Class I device recall on Jun 22, 2022 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class I
- Started: Jun 22, 2022
- Status: Ongoing
Integra has received complaints associated with ICP readings drifting to -50 mmHg (out-of-range) and may manifest an error message "sensor or extension cable failure!" The firm has found that this is caused by electrical interference from a component of the monitor's circuit board and from the environment.
- Product
- CereLink ICP Monitor; Catalog No. 826820, 826820P. Intracranial pressure monitor.
- Quantity
- 1,210 units
- Lots and codes
- UDI-DI- 10381780533778; All serial numbers.
- Distribution
- Worldwide Distribution: US nationwide, Australia, New Zealand, Canada, Hong Kong, Pakistan, Georgia, Saudi Arabia, Israel, South Africa, United Arab Emirates, Austria, Belgium, Czech Republic, Denmark, Estonia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Latvia, Lithuania, Netherlands, Norway, Portugal, Switzerland, Spain, Sweden, Slovakia, and United Kingdom.
- Recalling firm
- Integra LifeSciences Corp., Princeton, NJ, United States
- How the firm notified
- Letter
- FDA classified
- Jul 22, 2022
Newest first, one product per recall event; search them all
- AURORA Surgiscope System, Sterile, single use device that contains a Sheath, Obturator, and Imager. Manufacturer's…Z-1295-2025 · Class II · Feb 6, 2025 · Ongoing
- Brand Name: MediHoney¿ Gel Product Name: MediHoney¿ Wound & Burn Dressing Gel Model/Catalog Number: 31815 Software…Z-1168-2025 · Class II · Jan 10, 2025 · Ongoing
- CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810PL. Implantable device that…Z-0978-2025 · Class II · Dec 16, 2024 · Ongoing
- Duraform Dural Graft Implant, Size 3x3 , Model Number 80-1477US Lot CT007694Z-0262-2025 · Class II · Sep 26, 2024 · Ongoing
- SURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated for the…Z-0240-2025 · Class II · Sep 20, 2024 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports