Digital VIPR ROC-9839 Roberts Oxygen Ready to Use Digital VIPR (head only) ROC-9840
Recall Z-1376-2023, part of FDA recall event 91561. See Western/Scott Fetzer's company page.
Western/Scott Fetzer Company started this Class II device recall on Jan 3, 2023 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Jan 3, 2023
- Status: Ongoing
Gauge malfunction: Digital gauge displays an incorrect tank pressure - displays pressure higher than actual tank pressure. This may lead to tank running out of oxygen during use without audible alarms, potentially leading to an out of oxygen event and no flow.
- Product
- Digital VIPR ROC-9839 Roberts Oxygen Ready to Use Digital VIPR (head only) ROC-9840, Roberts Oxygen Ready to Use Digital VIPR System (with E-size cylinder) MNDR-600, Western Oxytote DTE Digital VIPR (head only) MNDS-602, Western Oxytote DTE Digital VIPR System (with D-size cylinder) MNDS-603, Western Oxytote DTE Digital VIPR System (with E-size cylinder)
- Quantity
- 24,287 units
- Lots and codes
- ROC-9839, Roberts Oxygen Ready to Use Digital VIPR (head only) o UDI-DI: B929ROC98390 ROC-9840, Roberts Oxygen Ready to Use Digital VIPR System (with E-size cylinder) o UDI-DI: B929ROC98400 MNDR-600, Western Oxytote DTE Digital VIPR (head only) o UDI-DI: B929MNDR6000 MNDS-602, Western Oxytote DTE Digital VIPR System (with D-size cylinder) o UDI-DI: B929MNDS6020 MNDS-603, Western Oxytote DTE Digital VIPR System (with E-size cylinder) o UDI-DI: B929MNDS6030 All units distributed from 05/29/2020 to 12/7/2022.
- Distribution
- US Nationwide distribution in the states of CO, CT, ID, MD, MT, TN, WA.
- Recalling firm
- Western/Scott Fetzer Company, Westlake, OH, United States
- How the firm notified
- Letter
- FDA classified
- Apr 12, 2023