Foot Controls Model: S-N1 and S-N2 Product Usage: Variable foot control for use with…
Recall Z-1351-2018, part of FDA recall event 79533. See W & H DentalWerk Burmoos GMBH - Site 1's company page.
W & H DentalWerk Burmoos GMBH - Site 1 started this Class II device recall on Dec 21, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated Jun 7, 2019.
- Classification: Class II
- Started: Dec 21, 2017
- Status: Terminated
During continuous postmarket surveillance activities, we have noticed a potential malfunction of individual foot controls of the type S-N1/S-N2. In case the error occurs, the foot control might not switch back to zero position when the pedal is released. Consequently, the motor keeps rotating. If this happens during automatic thread cutting or automatic screwing-in of implants, it may, in the worst case, lead to a serious injury of the patient.
- Product
- Foot Controls Model: S-N1 and S-N2 Product Usage: Variable foot control for use with drive units operating mechanical instruments.
- Quantity
- —
- Lots and codes
- REF: 05046200 05046200 05046211 05046210 06202400 06202400 30285000 30285001 Serial number "SN" from to (including) 22400 23429 E02201 E02485 E01011 E01033 E01035 E01071 24315 34040 E01358 E01658 01001 02318
- Distribution
- US Nationwide Distribution
- Recalling firm
- W & H DentalWerk Burmoos GMBH - Site 1, Burmoos, Austria
- How the firm notified
- Letter
- FDA classified
- Apr 10, 2018
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports