Arca
Skip to content
Tools
Tools

VITROS Immunodiagnostic Products TSH Reagent Pack

Recall Z-1322-2018, part of FDA recall event 79508. See Clinical Diagnostic Systems's company page.

Clinical Diagnostic Systems started this Class III device recall on Jan 11, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated Sep 8, 2020.

  • Classification: Class III
  • Started: Jan 11, 2018
  • Status: Terminated

Why It Was Recalled

Class III

VITROS TSH reagent lots have been found to generate higher than expected Calibrator level 2 signals on some customer systems. These Calibrator Level 2 responses cause a higher than expected frequency of customer generated calibration curves to fall outside of the calibration quality parameter measuring the gradient of the Customer Calibration Curve between Calibrator Level 2 and Calibrator Level 3.

Record

Product
VITROS Immunodiagnostic Products TSH Reagent Pack
Quantity
Lot #5470 (US=8570/OUS=537); Lot #5430 (US=0/OUS=9135); Lot #5178 (US=5178/OUS=4396)
Lots and codes
Lot 5470, expiry date: 24-May-2018 Lot 5430, expiry date: 01-Mar-2018 (not distributed in the US.) Lot 5178 was also affected by this issue, expired on 31-Mar-2017,
Distribution
Distributors were notified via letter. Lot 5430 was not distributed in the US. Lot 5178 was also affected by this issue, but this lot expired on 31-Mar-2017 so Ortho took no action on this lot.
Recalling firm
Clinical Diagnostic Systems, Rochester, NY, United States
How the firm notified
—
FDA classified
Apr 11, 2018

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports