AOA Distalizer Lock Nut Screw
Recall Z-1314-2014, part of FDA recall event 67647. See Allesee Orthodontic Appliances's company page.
Allesee Orthodontic Appliances started this Class II device recall on Feb 24, 2014 (voluntary: firm initiated); FDA lists it as terminated, terminated Aug 23, 2014.
- Classification: Class II
- Started: Feb 24, 2014
- Status: Terminated
The Lock Nut component of the Jet appliances may be missing the screw collar, which serves as a mechanical stop to prevent over-tightening in appliance assemby. Without this collar, an over-tightened lock nut may stop the piston assembly from moving freely, subsequently preventing intended movement of the teeth.
- Product
- AOA Distalizer Lock Nut Screw, found in products listed as: Distal Jet, Spring Jet, Mesial Jet, and Uprighter Jet. Removable functional orthodontic appliances that are to modify the growth of the jaws in a prescribed growth pattern.
- Quantity
- 411
- Lots and codes
- All Devices/Codes Manufactured from August 28, 2013 to November 6, 2013
- Distribution
- Worldwide Distribution: US (nationwide) and the country of : Canada.
- Recalling firm
- Allesee Orthodontic Appliances, Sturtevant, WI, United States
- How the firm notified
- Letter
- FDA classified
- Apr 1, 2014
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports