TISSUE APPROXIMATION SYSTEM (TAS)
Recall Z-1294-2025, part of FDA recall event 96274. See TAS Medical's company page.
TAS Medical Inc started this Class II device recall on Jan 27, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Jan 27, 2025
- Status: Ongoing
Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
- Product
- TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP
- Quantity
- 114
- Lots and codes
- Model/UDI-DI/Lot(Expiration): T-4000/00197644217267/230403(2024/04/20); T-5000/00860011435921/240801(2025/08/28), 240901(2025/09/25); T-LAP/00860011435914, 00197644927241/240702(2025/07/08), 230906(2024/11/29), 230905(2024/11/29)
- Distribution
- US Nationwide distribution in the states of LA, OK, TN.
- Recalling firm
- TAS Medical Inc, Windermere, FL, United States
- How the firm notified
- Letter
- FDA classified
- Mar 6, 2025
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports