Arca
Skip to content
Tools
Tools

Philips Connected Blood Glucose Meter

Recall Z-1270-2025, part of FDA recall event 96230. See Telcare's company page.

Telcare, LLC started this Class II device recall on Jan 30, 2025 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Jan 30, 2025
  • Status: Ongoing

Why It Was Recalled

Class II

Telcare, LLC has received customer complaints related to battery expansion/swelling for its Philips Connected Blood Glucose Meter (i.e., BGM 4 brand name) Monitoring System.. The issue was identified by post market surveillance data (customer complaints) - events of battery expansion/swelling and associated symptoms, e.g., leading to the device's removable back cover (battery port) to separate. If there is a loss of primary function of the device due to the described device defect, an immediate health consequence that may result is prolonged hypoglycemia or hyperglycemia due to delay in treatment, as the blood glucose measurement is not available to guide treatment. In a worst-case situation, there is potential for battery swelling to result in explosion of the battery and/ or device. Exposure to explosion, and subsequently projectiles from the explosion, may result in several immediate health consequences, including concussion, headache, dizziness, eye irritation, blurred vision, eye injury, and vision impairment.

Record

Product
Philips Connected Blood Glucose Meter, BGM 4; Model Number: TM0009;
Quantity
34,764 units
Lots and codes
Model Number: TM0009; UDI-DI: 00859519002407; Serial Numbers: Serial Numbers are formatted as TFXXXXXX or TGXXXXXX (X represents digits). Meter Serial Numbers: TF048802, TF098191, TF079123, TF076786, TF076836, TF089080, TF088984, TF086097, TF085390, TF085029, TF084238, TF098751, TF098975, TF098976, TF096704, TF096961, TF096966, TF117862, TF127077, TF049583, TF047153, TF063135, TF127267, TF100376, TF086287, TF104396, TF104399, TF104401, TF104403, TF104404, TF104405, TF104407, TF104408, TF104409, TF104410, TF104411, TF104412, TF104414, TF104415, TF104416, TF104417, TF104418, TF104421, TF104426,…
Distribution
US Nationwide distribution.
Recalling firm
Telcare, LLC, Concord, MA, United States
How the firm notified
Letter
FDA classified
Mar 5, 2025