human med Biofill Infiltration Cannula
Recall Z-1251-2025, part of FDA recall event 96137. See HUMAN MED's company page.
HUMAN MED AG started this Class II device recall on Oct 18, 2024 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Oct 18, 2024
- Status: Ongoing
The cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather than a fan-shaped spray jet.
- Product
- human med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use.
- Quantity
- 775 devices
- Lots and codes
- Lot number 24-20198, UDI (01)04260170881014(11)240611(17)270611(10)24-20198; and Lot number 23-19612, UDI (01)04260170881014(11)240109(17)270109(10)23-19612.
- Distribution
- US Nationwide distribution in the state of FL.
- Recalling firm
- HUMAN MED AG, Schwerin, Germany
- How the firm notified
- Letter
- FDA classified
- Feb 28, 2025
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports