DxC 500 AU Clinical Chemistry Analyzer

Recall Z-1233-2025, part of FDA recall event 96193. See Beckman Coulter Laboratory Systems (Suzhou)'s company page.

BECKMAN COULTER LABORATORY SYSTEMS (SUZHOU) CO., LTD. started this Class II device recall on Jan 10, 2025 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Jan 10, 2025
  • Status: Ongoing

Why It Was Recalled

Class II

Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.

Record

Product
DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.
Quantity
56
Lots and codes
Software Versions: V1.4, V1.4.1, V1.4.2, V1.4.3 REF/UDI-DI/Serial Numbers: C63519/14987666545058/2024090170; C63520/14987666545065/2024070116, 2023070009, 2024010038, 2024010035, 2024050072, 2024040056, 2024030049, 2024020046, 2024070111, 2024030051, 2024030052, 2024010033, 2023090015, 2024060103, 2024060104, 2024060101, 2024080145, 2024020045, 2024020042, 2024020043, 2024020047, 2024060096, 2024040059, 2023100019, 2024030053, 2023070010, 2024070129, 2024070130, 2024020041, 2024090183, 2024010036, 2024010037, 2024070132, 2024070131, 2023100020, 2024050064, 2024050070, 2023090016, 2023110022,…
Distribution
US: MA, CA, SC, LA, OK, GA, AL, TX, OR, ID, CO, MN, NY, MS, PA, WA, IL, MI, PR. OUS: Lebanon, Poland, Ghana, Spain, South Africa, Slovakia, Ghana, Croatia, Seychelles, Australia, Republic of Korea
Recalling firm
BECKMAN COULTER LABORATORY SYSTEMS (SUZHOU) CO., LTD., Suzhou, N/A, China
How the firm notified
Letter
FDA classified
Feb 25, 2025