Arca
Skip to content
Tools
Tools

Presource PBDS, Knee Arthroscopy

Recall Z-1197-2013, part of FDA recall event 64736. See Cardinal Health Medical Products and Services's company page.

Cardinal Health, Medical Products & Services started this Class I device recall on Mar 26, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 12, 2017.

  • Classification: Class I
  • Started: Mar 26, 2013
  • Status: Terminated
  • Products in this event: 14

Why It Was Recalled

Class I

Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow.

Record

Product
Presource PBDS, Knee Arthroscopy, Kit, Circulator
Quantity
4,571 Kits total
Lots and codes
Catalog Number: PO24AKNPC01; Lots: 851288, 877594, 901672, 917786, 948443, 972424
Distribution
Distributed in the states of AR, MS, SC, and TN.
Recalling firm
Cardinal Health, Medical Products & Services, Park City, IL, United States
How the firm notified
Letter
FDA classified
May 3, 2013

Other Products in This Recall

Recall event 64736

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.