Arca
Skip to content
Tools
Tools

Cannabinoids DAU Calibrator Control Level 1 (37.5 ng/mL) This solution is prepared from…

Recall Z-1178-2014, part of FDA recall event 67521. See Lin-Zhi International's company page.

Lin-Zhi International Inc started this Class II device recall on Feb 6, 2014 (voluntary: firm initiated); FDA lists it as terminated, terminated Mar 10, 2015.

  • Classification: Class II
  • Started: Feb 6, 2014
  • Status: Terminated
  • Products in this event: 2

Why It Was Recalled

Class II

Product catalog #s 0007c Cannabinoid 37.5 ng/mL Control Level 1 and 0008c Cannabinoid 62.5 Control Level 2 of the lots mentioned may give decreased concentration readings towards the end of their shelf life due to use of dated but unexpired secondary stock used to create the products.

Record

Product
Cannabinoids DAU Calibrator Control Level 1 (37.5 ng/mL) This solution is prepared from non-sterile urine and should be handled as potentially infectious. IVD; The Cannabinoids (cTHC also known as THC version c) DAU negative and positive controls are human urine-based liquids and ready to use. The constituent is a processed drug-free human urine matrix, containing buffers, stabilizers, and less than 0.09 % of sodium azide. The controls are prepared by spiking known concentrations of 11-nor-9-THC-9-COOH into the drug-free matrix. cTHC is referenced as an abbreviation for 11-nor-9-THC-9-COOH and it may be referred to simply as THC in the text of this report. The controls contain 37.5 ng/mL and 62.5 ng/mL of cTHC,¿ 10% by GC/MS or LC/MS analysis. Lin Zhi International inc. Sunnyvale, CA 94085 The Cannabinoid Drug of Abuse (DAU) negative and positive controls are used as assay quality control material to monitor the precision of the Lin-Zhi International, Inc. (LZI) Cannabinoid Enzyme Immunoassay (Ref # 0070c) on a number of automated clinical chemistry analyzers.
Quantity
130 items
Lots and codes
Catalog number 0007Cm; lot number 1211158R; Expiration May 14, 2014.
Distribution
US Distribution including the states of SC, NC, LA, FL, TX,, AL and GA and Internationally to Germany.
Recalling firm
Lin-Zhi International Inc, Sunnyvale, CA, United States
How the firm notified
E-Mail
FDA classified
Mar 12, 2014

Other Products in This Recall