Arca
Skip to content
Tools
Tools

ClariVein IC infusion catheter

Recall Z-1146-2018, part of FDA recall event 79324. See Vascular Insights's company page.

Vascular Insights, LLC started this Class II device recall on Nov 9, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated Dec 18, 2018.

  • Classification: Class II
  • Started: Nov 9, 2017
  • Status: Terminated

Why It Was Recalled

Class II

Certain catheters could have a compromised catheter shaft due to a manufacturing defect which could result in some delamination of the catheter shaft including during introduction of the catheter into a vascular access introducer.

Record

Product
ClariVein IC infusion catheter, Model Number: 65-018-E4S Product Usage: The ClariVein IC infusion catheter is indicated for infusion of physician-specified agents into the peripheral vasculature.
Quantity
928
Lots and codes
Lot number: BM651403017I
Distribution
U.S Nationwide in the states: AZ, CA, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, MO, NE, NJ, NY, OH, PA, SC, TN, TX, VA, WA
Recalling firm
Vascular Insights, LLC, Quincy, MA, United States
How the firm notified
E-Mail
FDA classified
Mar 20, 2018

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports