Denlase: Diode Laser Therapy System and Penlase: Dental Laser Therapy System surgical…
Recall Z-1145-2014, part of FDA recall event 67520. See China Daheng Group's company page.
China Daheng Group Inc. started this Class II device recall on Jul 15, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 17, 2015.
- Classification: Class II
- Started: Jul 15, 2013
- Status: Terminated
Failure to provide calibration procedures to purchasers, as required by 21 CFR 1040.11(a)(2). For example, the user manual did not contain calibration procedures for Denlase or Penlase Lasers
- Product
- Denlase: Diode Laser Therapy System and Penlase: Dental Laser Therapy System surgical powered. Surgical powered Laser instruments.
- Quantity
- 238
- Lots and codes
- Denlase and Penlase Dental Laser Systems.
- Distribution
- CA, KS, IL, PA, NJ, WI, WY, KY, FL and NY India, Indonesia, Turkey, Denmark, and France,
- Recalling firm
- China Daheng Group Inc., Beijing, China
- How the firm notified
- Visit
- FDA classified
- Mar 14, 2014
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports