Arca
Skip to content
Tools
Tools

Denlase: Diode Laser Therapy System and Penlase: Dental Laser Therapy System surgical…

Recall Z-1145-2014, part of FDA recall event 67520. See China Daheng Group's company page.

China Daheng Group Inc. started this Class II device recall on Jul 15, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 17, 2015.

  • Classification: Class II
  • Started: Jul 15, 2013
  • Status: Terminated

Why It Was Recalled

Class II

Failure to provide calibration procedures to purchasers, as required by 21 CFR 1040.11(a)(2). For example, the user manual did not contain calibration procedures for Denlase or Penlase Lasers

Record

Product
Denlase: Diode Laser Therapy System and Penlase: Dental Laser Therapy System surgical powered. Surgical powered Laser instruments.
Quantity
238
Lots and codes
Denlase and Penlase Dental Laser Systems.
Distribution
CA, KS, IL, PA, NJ, WI, WY, KY, FL and NY India, Indonesia, Turkey, Denmark, and France,
Recalling firm
China Daheng Group Inc., Beijing, China
How the firm notified
Visit
FDA classified
Mar 14, 2014

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports