MAVERICK Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN…
Recall Z-1024-2026, part of FDA recall event 98260. See Repligen's company page.
REPLIGEN CORPORATION started this Class II device recall on Dec 10, 2025 (fda mandated); FDA lists it as ongoing.
- Classification: Class II
- Started: Dec 10, 2025
- Status: Ongoing
Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
- Product
- MAVERICK Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
- Quantity
- 33
- Lots and codes
- PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
- Distribution
- U.S.
- Recalling firm
- REPLIGEN CORPORATION, Marlborough, MA, United States
- How the firm notified
- —
- FDA classified
- Jan 13, 2026
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports