Intraosseous Fixation System (IFS) 1.7 mm

Recall Z-0975-2013, part of FDA recall event 64485. See OrthoHelix Surgical Designs's company page.

Orthohelix Surgical Designs Inc started this Class II device recall on Oct 30, 2012 (voluntary: firm initiated); FDA lists it as terminated, terminated Dec 5, 2013.

  • Classification: Class II
  • Started: Oct 30, 2012
  • Status: Terminated

Why It Was Recalled

Class II

The firm was notified by a customer that there was a burr on the inside end of the drill bit resulting in the drill bit not fitting over the associated guide wire.

Record

Product
Intraosseous Fixation System (IFS) 1.7 mm, 2.0 mm & 2.5 mm Cannulated Drill Bit. Intended to stabilize and aid in fixation of fractures.
Quantity
Lot # 2399041, 97 pcs; Lot # 2399051, 143 pcs; Lot # 2399051R, 4 pcs & Lot # 2399061, 119 pcs
Lots and codes
Part #: IFS-072-17-C, Lot #: 2399041; Part #: IFS-72-20-C, Lot #: 2399051 & Lot #: 2399051R and Part #: IFS-072-25-C, Lot #: 2399061.
Distribution
Nationwide Distribution including the states of CA, CO, IL, MI, MD, NC, NM, OH, OK, PA, SD, TX, UT, VA, WA and WI.
Recalling firm
Orthohelix Surgical Designs Inc, Medina, OH, United States
How the firm notified
Letter
FDA classified
Mar 22, 2013