GOPump Elastometric infusion PumpKit with/ and without accesssories all containing Dual…
Recall Z-0940-2013, part of FDA recall event 64423. See Symbios Medical Products's company page.
Symbios Medical Products, LLC started this Class I device recall on Feb 25, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Feb 13, 2014.
- Classification: Class I
- Started: Feb 25, 2013
- Status: Terminated
Received 2 complaints out of 458 distributed pumps in this lot, where flow restrictor bead became displaced, which were reported to permit fast flow of contents.
- Product
- GOPump Elastometric infusion PumpKit with/ and without accesssories all containing Dual GOPump, part number 510076: One infusion pump comprising two, 150 mL elastomeric reservoirs, each of which connects to its own catheter. Liquid flow rate is labeled at 2 mL/hr per elastomer. The Symbios Disposable Infusion Pump Kit is a disposable, self-contained infusion system utilizing an inflatable elastomeric reservoir to mechanically provide percutaneous infusion of prescribed solutions at a pre-set rate for post-operative pain management
- Quantity
- 458
- Lots and codes
- Kit Part Number 510545-BP, lot 12-100983, 12-100984, 12-101096, 12-101125 Kit Part Number 510112-BP, lot 12-101110, 12-101103 Kit: 510401-BP lot 12-101045 Kit: 510644, lot 12-101028, 12-101097, 12-101130 GOPump Dual Infusion Pump Kit All Kits contain Pump Part Number: 510076 Lot Number: 12-100601
- Distribution
- Nationwide Distribution including PA, CA, OH, AL, FL, WV, TN, NJ, NC, TX, CO, NY, TX, MO, MS, SC, and GA.
- Recalling firm
- Symbios Medical Products, LLC, Indianapolis, IN, United States
- How the firm notified
- Letter
- FDA classified
- Mar 13, 2013
Newest first, one product per recall event; search them all
- Part 510450 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 10mL/hr ),... found in the following kits GoBlock…Z-1458-2013 · Class I · May 10, 2013 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports