Arca
Skip to content
Tools
Tools

DIASTAT anti-Nuclear antibody (ANA) / DIASTAT ANA ELISA. Catalog Number: FANA200. Product…

Recall Z-0929-2018, part of FDA recall event 79183. See Euro-Diagnostica's company page.

Euro Diagnostica AB started this Class II device recall on Oct 15, 2015 (voluntary: firm initiated); FDA lists it as terminated, terminated Mar 29, 2019.

  • Classification: Class II
  • Started: Oct 15, 2015
  • Status: Terminated

Why It Was Recalled

Class II

Complaint investigation concluded that although product quality requirements stated in the instructions for use were met at lot release, the high background in the ELISA plate, the strong IgM conjugate, and the low reference control led to an increased risk of equivocal and/or false positive sample test results for the two FANA200 kit lots SS 1500 and SS 2009.

Record

Product
DIASTAT anti-Nuclear antibody (ANA) / DIASTAT ANA ELISA. Catalog Number: FANA200. Product Usage: The DIASTAT anti-nuclear antibody (ANA) test is a qualitative enzyme-linked immunosorbent assay (ELISA) for the detection of ANAs in human serum or EDTA plasma. It detects ANAs against Sm, Sm/RNP, Ro (SS-A), La (SS-B), Scl- 70, Jo-1, dsDNA, histone and centromere antigens.The test may be used to screen out samples which are negative for all ANAs. Samples that give a positive test result should be further tested to identify the antigen-specific antibody or antibodies present. Quantitative/qualitative ELISA kits are available for the individual detection of Sm, Sm/RNP, Ro (SS-A), La (SS-B), dsDNA and centromere ANAs. Qualitative kits are available for the individual detection of Scl-70 and Jo-1. ANA detection represents one parameter in a multicriterion diagnostic process.
Quantity
1090
Lots and codes
UDI: 07340058410486. Lot numbers: SS 1500 and SS 2009.
Distribution
Worldwide Distribution and US. Nationwide
Recalling firm
Euro Diagnostica AB, Malmo, N/A, Sweden
How the firm notified
E-Mail
FDA classified
Mar 8, 2018

Euro Diagnostica AB's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports