Giraffe Blue Spot PT LiteTM Phototherapy System
Recall Z-0923-2020, part of FDA recall event 84562. See Lumitex's company page.
Lumitex Inc started this Class II device recall on Dec 11, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Sep 8, 2021.
- Classification: Class II
- Started: Dec 11, 2019
- Status: Terminated
There is a potential for some systems built between 2014 and Jan 2018 to have a light output falling below the recommended minimum output. This problem has been associated with exposure of the light pipe to elevated temperatures, which is contra-indicated in the Operation, Maintenance, and Service Manual.
- Product
- Giraffe Blue Spot PT LiteTM Phototherapy System, Model Numbers M1224917, M1231533, M1225025, M1224961, M1224958, M1224957, M1224959, M1224963, M1225773, M1225066, M1224960, M1225059, M1225023, M1225060, M1225063, M1225064, M1225024, M1231622, M1236303, M1238489, M1238490, M1238492, M1238493, M1238495, M1238499, M1238503, M1238505, M1238507, M1238508, 2063860-001, 2063862-001, 2063870-001, 2063873-001, 2063874-001, 2071393-001, 2080916-001, 2089424-001
- Quantity
- 6837
- Lots and codes
- 2014 Production Serial Numbers: All Serial numbers beginning with QAAT; 2015 Production Serial Numbers: All Serial numbers beginning with QAAU; 2016 Production Serial Numbers: All Serial numbers beginning with QAAV; 2017 Production Serial Numbers: All Serial numbers beginning with QAAW; 2018 Production Serial Numbers: Serial numbers QAAT60001 through QAAT60156
- Distribution
- The products were distributed to the following US states: WI.
- Recalling firm
- Lumitex Inc, Strongsville, OH, United States
- How the firm notified
- Letter
- FDA classified
- Jan 31, 2020
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports