nuwellis AquaFlexFlow UF 500 Plus
Recall Z-0831-2025, part of FDA recall event 95991. See Nuwellis's company page.
Nuwellis Inc started this Class I device recall on Dec 11, 2024 (voluntary: firm initiated); FDA lists it as completed.
- Classification: Class I
- Started: Dec 11, 2024
- Status: Completed
The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.
- Product
- nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extracorporeal blood circuit which is used with the Aquadex SmartFlow or Aquadex FlexFlow System
- Quantity
- 845 units
- Lots and codes
- Lot Numbers: 22697, 22698, 22699, 22700, 22701, 22702, 22703, 22704, 22705, 22706, 22707, 22708, 22721, 22722, 22723, 22724, 22733, 22734, 22735, 22736, 22737, 22738, 22739, 22740.
- Distribution
- US Nationwide Distribution and the country of Hong Kong
- Recalling firm
- Nuwellis Inc, Eden Prairie, MN, United States
- How the firm notified
- FDA classified
- Jan 16, 2025
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports