Arca
Skip to content
Tools
Tools

TactiFlex Ablation Sensor Enabled Ablation Catheter

Recall Z-0814-2024, part of FDA recall event 93711. See St. Jude Medical Atrial Fibrillation Division's company page.

St. Jude Medical, Atrial Fibrillation Division, Inc. started this Class II device recall on Dec 18, 2023 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Dec 18, 2023
  • Status: Ongoing

Why It Was Recalled

Class II

When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific to the catheter do not appear on the display and the catheter image may also appear inverted.

Record

Product
TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile.
Quantity
483 units
Lots and codes
Lot numbers 10073391, 10075618, 10080798, 10086492, 10087953, 10089624, 10092157, 10093066, and 10095363, Exp. 11/30/2024, GTIN 05415067034618.
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DC, DE, FL, GA, IA, IL, KS, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, and WV. The country of Canada.
Recalling firm
St. Jude Medical, Atrial Fibrillation Division, Inc., Saint Paul, MN, United States
How the firm notified
Letter
FDA classified
Jan 25, 2024

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.