Arca
Skip to content
Tools
Tools

SAM CHEST SEAL REF CS203-EN Combo LOT Y060321-09

Recall Z-0789-2022, part of FDA recall event 89510. See Sam Medical Products's company page.

Sam Medical Products started this Class II device recall on Jan 28, 2022 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 6, 2023.

  • Classification: Class II
  • Started: Jan 28, 2022
  • Status: Terminated

Why It Was Recalled

Class II

Due to partially assembled valved dressing that has a hole within it instead of the intended non-valved dressing in the package resulting in that the dressing cannot be centered over the open chest wound and create an occlusive seal due to the hole being present.

Record

Product
SAM CHEST SEAL REF CS203-EN Combo LOT Y060321-09
Quantity
2,950 packages
Lots and codes
Model Number: CS203-EN UDI Code: (01)10822045000190(10)Y060321-09(17)270601 (Case Carton) UDI Code: (01)10822045000193(10)Y060321-09(17)270601 (Each Pouch/Package) Lot Number: Y060321-09
Distribution
U.S.: AZ, CA, CO, FL, GA, MA, NY, OR, PA, Puerto Rico, RI, TX, VA, and WA O.U.S.: Canada
Recalling firm
Sam Medical Products, Tualatin, OR, United States
How the firm notified
Letter
FDA classified
Mar 16, 2022

Sam Medical Products's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports