Arca
Skip to content
Tools
Tools

Endobon-Xenograft Granules

Recall Z-0788-2016, part of FDA recall event 73141. See Biomet France's company page.

Biomet France SARL started this Class II device recall on Jan 21, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated May 31, 2017.

  • Classification: Class II
  • Started: Jan 21, 2016
  • Status: Terminated

Why It Was Recalled

Class II

Cytotoxicity tests were not passed at 36 months. Use of Endobon-Xenograft Granules past 18 months of its specified shelf life may result in patient illness/injury.

Record

Product
Endobon-Xenograft Granules, Zimmer Biomet, Synthetic Bone Grafting Material, used in dental and/or surgical procedures. Endobon Xenograft Granules, Bone Grafting Material, Synthetic are used in the following dental and/or surgical procedures: " Alveolar ridge augmentation/reconstruction, " Sinus elevation, " Filling of resection defects in benign bone tumor, bone cysts, or other defects in the alveolar ridge or wall, " Filling of bone defects after apicectomy, " Filling alveoli after tooth extraction. This product should not be used in non-peridontal mandibular applications.
Quantity
145,116 units
Lots and codes
Item Numbers &Lot Numbers: ROX05 S016303 ROX05 S016304 ROX05 S0163056 ROX05 S0163059 ROX05 S0171051 ROX05 S017105R ROX05 S0171098 ROX05 T0001122 ROX05 T0001133 ROX05 T0001136 ROX05 T0001141 ROX05 T0001147 ROX05 T0025073 ROX05 T0034084 ROX05 T0045150 ROX05 T0045157 ROX05 T0052108 ROX05 T0064164 ROX05 T0064168 ROX05 T0064175 ROX05 T0096191 ROX05 T0104217 ROX05 T0104238 ROX05 T0114238 ROX05 T0122226 ROX05 T0122238 ROX05 T0134261 ROX05 T0134269 ROX05 T0142245 ROX05 T0142248 ROX05 T0162294 ROX05 T0183280 ROX05 T0192311 ROX05 T0192316 ROX05 T0205297 ROX05 T0214319 ROX05 T0224329 ROX05 T0224333…
Distribution
Worldwide Distribution - US including AL, AK, AR, AK, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MN, MI, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV and Internationally to Canada, Japan, Europe.
Recalling firm
Biomet France SARL, Valence Cedex 9, N/A, France
How the firm notified
Letter
FDA classified
Feb 9, 2016

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports