NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi…
Recall Z-0773-2019, part of FDA recall event 81920. See Neusoft Medical Systems Imp & Ex's company page.
NEUSOFT MEDICAL SYSTEMS IMP & EX started this Class II device recall on Jan 9, 2019 (fda mandated); FDA lists it as terminated, terminated Aug 25, 2022.
- Classification: Class II
- Started: Jan 9, 2019
- Status: Terminated
For surview scan length more than 500mm,if the user aborts or skips the reconstructed surview, the image may have the wrong field of view (FOV). When user plans and performs subsequent clinical scan with the wrong FOV the actual start and end position may not be consistent with the planned positons.
- Product
- NeuViz 128 Multi-slice CT Scanner System with software version 1.0.8.5010O+P04 Vi .0.8.521 9+PO1 Multi-Slice CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array. The acquired X-RAY transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.
- Quantity
- 3
- Lots and codes
- Series number includes: N1281180242E, N1281170231E, N1281180248E
- Distribution
- US Distribution in states of: KS, OK and MA.
- Recalling firm
- NEUSOFT MEDICAL SYSTEMS IMP & EX, Shenyang, China
- How the firm notified
- Letter
- FDA classified
- Feb 5, 2019
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports